The Vendor is required to provide to procure a nanoparticle tracking analysis tool include:
• Nanoparticle tracking analyzer for analyzing size, concentration and zeta potential in fluorescence and scatter mode.
• At least 4 fluorescence channels.
• Excitation at 405nm, 488nm, 520nm and 640nm.
• All excitation sources can be used on the same sample.
• Size range: 20-1000nm.
• Concentration range: 10^5- 10^9 particles/ml.
• Analysis of simultaneous expression of two markers on one particle.
• Analyzed volume should be bigger than 20nl for sufficient statistics.
• Time for one analysis should be below 1min.
• No use of fluidic pumps during analysis to avoid impact of liquid flow on results.
• Live size and concentration tracking.
- Contract Period/Term: 3 years