A government agency in British Columbia issued “Automated Liquid Handling System” (RFP). Solicitation SYS-3548 closed on January 6, 2026. Eligibility: Only for Canada Organization.
The vendor required to provide automated liquid handling system, that will be used to transiently transfect mammalian cells in 24- and 96-well plate-based formats and subsequently incubate for protein expression.
- This system will be one of three components required for this suite:
• Automated liquid handling system
• A plate-based cell counter
• Biological safety cabinet
- A cell counter and incubator will be located on a standard laboratory bench beside the agency.
- This automated liquid handling system should be provided as a separate stand-alone piece.
- The automated liquid handling system required for transfect mammalian cells in a plate-based format.
- The system should be flexible with the plate format, either 24- or 96-well, microplate or deep-well plate.
- The equipment should be compatible with different transfection protocols and capable of handling volumes from 0.1 – 4 ml.
- This system should have a minimum of 8 independent 1-1000 µl pipetting channels and 4 independent 5 ml pipetting channels.
- The system should contain a multifunctional deck space that is the at least the size of 20 microplates.
- The deck should be flexible and configurable to accommodate multiple different set ups depending on plate, tip, tube and vial needs.
- There will be situations when the system will need to transfer to and from at least 12x50 ml conical tubes or at least 24x15 ml conical tubes.
- The deck space should be able modifiable to manage either at least 12x50 ml conical tubes or at least 24x15 ml conical tubes.
- the system should be able to move either microplates or deep-well plates to and from any location on the deck space.
- The system should be able to remove plate lids or hold them during pipetting events.
- The system should be able to hold at least 5 stacks of tip racks.
- The system should contain a waste module for liquid waste.
- The system should contain a tilt module which can tilt plate at least 10° to ease the removal of used media.
- The system should contain at least 6 reagent reservoirs at 60 ml.
- The system should have holder for up to thirty-two 1.5/2 ml microcentrifuge tubes.
- This system should have temperature-controlled plate shaker to ensure cells maintain correct temperatures and remain suspended during the transfection protocol.
- The system should come with all components necessary for calibration and quality assurance checks.
- The system should be programmable by the user to develop in-house protocols.
- The system should include hands-on user training (on-site) for all functionality of the system and related software.
- Questions/Inquires Deadline: December 22, 2025
Ask any question about the RFP and its amendments. Instantly surface schedules, deliverables, compliance items, evaluation criteria, and budget details with 100% source-backed answers.
Quick answers about deadlines, eligibility and submission for this solicitation.
No. This opportunity (SYS-3548) closed on January 6, 2026. See similar open RFPs below.
A government agency in British Columbia. The buyer's contact and submission details are in the solicitation documents.
Only for Canada Organization.
Questions and inquiries must be submitted by December 22, 2025.
Yes. After purchase, with a subscription or during a free trial, the RFPPlanet AI Assistant answers questions sourced from this RFP’s documents, such as requirements, evaluation criteria and key dates.
Set up free email alerts and get notified when new government and private-sector bids and tenders match your industry and location. Choose daily or weekly delivery.